注册 忘记密码
稿件信息
稿件编号: L3827
稿件标题: 3款POCT血糖仪的性能分析
稿件作者: 邓济甦;
关键字词: 血糖仪;性能分析
杂志名称: 检验医学与临床
文章刊期: 2011年08卷04期
文章栏目: 论著
稿件作者(英文): dengjisu;
文章摘要(英文): (1 Clinical Laboratory, The First Affiliated Hospital Of Chongqing Medical University; Chongqing 400016,China) 【Abstract】Object:To evaluate the capability of different POCT glucose meter and understand the difference between the venous blood glucose results with POCT glucose meters and the blood glucose results tested by biochemical analyzer. Methods:The accuracy,reportable range,comparison with biochemical analyzer were detected by the different instrument with heparin whole blood. All statistical analyses were performed with SPSS version 11.0 statistical software.Results:The within-CV% of Sure-Step, Accu-Chek Active, ONE Touch POCT glucose meters were 2.36-2.54%、1.06-1.25%、4.20-4.66% respectively, regression equation with biochemical analyzer (AU5400) were Y = 1.0299X +0.1789, Y = 0.8611X +0.5276, Y = 0.9267X-0.0697 respectively.In the medical decision level,the differences from comparison with the target detection systems was <20%. Conclusion:There is significant differences between different POCT glucose meters and blood glucose and biochemical analysis of the results detected. POCT glucose testing method can be used as the selection, but not a substitute for blood glucose measured biochemical analyzer.
相关信息: 文件大小:1.18 MB
页数:2
浏览次数:1655
下载次数:211
免费查看全文 下载PDF阅读器